(CSVLAF, SMVLAF & HLAF)
It is explained as a booth for product sampling and dispensing. It is designed as an open-front containment system with a built–in scavenging arrangement that draws powder aerosol away from the operator and the operating environment, protecting the operator’s health and any product contamination.

GENERAL SPECIFICATION
| Sr. No. | Item | Description |
|---|---|---|
| 1. | MOC | SS 304/SS 316 / SS 316L combination of SS 304 & SS 316 as per – ASTMA |
| 2. | Three stage filtration | EU6 – FINE Filter (efficiency 95% down to 5 micron) EU14- HEPA Filter – Supply (99.999% down to 0.3 micron) Dynamically balanced with suitable rating and size |
| 3. | Motor Blower | LED Lights with milky white diffuser SS 304/Glass side panel 5/15 Amp switch socket for external equipment ON/OFF switch |
| 4. | Operating Conditions Air Cleanliness Air Velocity Air Flow Noise Level Vibration Level | ISO Class 5 (ISO 14644-1:2015) [E] 0.45 +-0.05 mps Vertical / Horizontal Less than 67 dB on scale ‘A’ when ambient is not greater than 55dB on scale ‘A’ Minimum |
| 5. | Optional Accessories | Vinyl Curtain Polycarbonate enclosures with Front Door (Hinged/ Pneumatic Cylinder) SS Perforated Safety Grill for HEPA Filter Audio Visual Alarm indicating condition of HEPA Filter U.V. Light with Hour meter Air/Gas Cock Vishwakarma Industries Data Management 21 CFR Part 11 compliance Software. |